KBM 80 plus TFT

Blood Bank Centrifuge

Blood Bank Centrifuge

KBM 80 Plus is the new generation microprocessor controlled blood bank centrifuge with forward thinking design and technology. Ideally used for centrifugation of whole blood for components separation like packed red cells, platelet rich plasma, platelet concentrate, cryoprecipitate and buffy coat etc.

 

Advance manufacturing techniques are used for mounting of brushless induction drive motor in PUF insulated, corrosion free, stainless steel armored chamber, resulting in minimized vibration and noise level. The smooth acceleration / deceleration helps in clear separation thus ensuring high quality component yield.

This centrifuge cater to the requirements of small, medium & large blood centers with flexibility to choose from 2 rotor options. The rotors are designed to accommodate specific type of oval shaped metal buckets & plastic carriers for holding eight &twelve blood bags of different volumes & configuration like 350 ml / 450m1, double/ triple/ quadruple, penta begs & inhne-filter (quintuple) blood bags.

User-friendly software & controls helps in ease of operation. There is a unique facility for setting & controlling the centrifuge run either by RPM (speed) or RCF forte) thus eliminating the need of manual calculations. Interactive menu driven program guides the operator through the setting process of run parameters & also the operational status while centrifuge is running. Specially designed wing-shielded rotors reduces friction which helps in energy saving & better temperature management.

KBM 80 Plus comes with microprocessor controlled advanced HMI Touch Screen

 

Features of Touch Screen

  • Facility to set 20 user profiles with identical password protected access
  • Protected admin control for key parameters
  • Display of realtime run graph with display set & run parameter
  • Gloves friendly touch screen operation with option to connect with wireless mouse
  • Onscreen manual, Troubleshooting & process guide

Performance

  • Speed holding accuracy +/- 10 RPM
  • Multiple acceleration & deceleration profiles
  • Temperature range from -20°C to +40°C
  • Low noise level (≤ 55db)
  • Choice of 2 wind-shielded rotors to process 8 & 12 blood bags per cycle
  • Tropicalized to run from 0°C to 40°C & RH up to 95%

Leading Tecnology

  • Brush-less induction motor with frequency drive, practically maintenance free
  • Advanced user friendly microprocessor control
  • Programmable centrifugation parameters for accurate control
  • Onboard real time user data log in with password protection
  • dedicated software for data analysis / PDF Reports, connectivity with USB
  • Auto-adjustable effective centrifugal force irrespective of rotor load

User Friendly

  • Facility to set and indicate RPM or RCF
  • Simultaneous display of set & run parameters
  • Self-diagnosis of program errors
  • Microprocessor controlled automatic lid opening with touch button
  • Log of 1000 run records on Touch screen
  • Display of real time & Date
  • Castor wheels with floor standing jacks for vibration free run
  • Emergency lid opening provision in case of power failure
  • Auto compressor cut off on lid opening

Safety

  • Automatic lid interlocking and unlocking
  • Imbalance cut-off with indication
  • Safety key lock to prevent unauthorized use
  • Rotor over speed protection
  • Alarms for imbalance, lid open, over temperature
  • Password protection for authorized access
  • Tamper proof memory & last program recall
  • Automatic Rotor identification & indication
  • Automatic recovery of process in case of power interruption ‘Catch On Fly”.

Easy Setup Programming & Monitoring

Technical Specifications
Parameter Unit KBM 80 Plus
Process controls Type HMI Touch screen
Menu driven program Nos 400
Acceleration profile Steps 1-15
Deceleration profile [incl. coasting] Steps 1-15
Max. speed RPM 4200 settable ±10 RPM
Max. RCF g 6300
Max time hh:min 0 to 99 hr 59 min 59 sec, Hold function
Compatible Rotors Code K 741e / K 751e
Dimension (W X D X H) mm 840 X 940 X 950
Compliance IEC 61010-2-020:2016, IEC 61010-1:2016, EN 60601-1, EN 60601-1-2, CE, US FDA

Optional: Servo controlled voltage stabilizer

Supply : 220-240 volts, 50Hz, Single phase

Blood Bank Rotors, Metal Buckets & Plastic Buckets

Data Acquisition System

These centrifuges have inbuilt system to collect & store all data related to centrifugation process like operator code, process type. time. temp. RPM. RCF. acceleration. deceleration & process status etc. The data from the centrifuge can be transferred to a computer using a data interface. This data can be used by blood centers for quality analysis S control.

REMI KBM 80 Plus meets the component separation guidelines by AABB & OCHS

Important Parameter
for Quality Components
Procedural
Requirements

(as per AABB & DGHS manual)*
Technical Features
KBM 80 Plus
Centrifugal Force (RCF) 5000 g
Heavy spin - Max. RCF
Up to 6300 g
Temperature Range Need temperature 22°C±2°C to run PRP
method & Buffycoat method, 4°C to run
FFP & Cryoprecipitate method
Achieves both desired temperature of
22°C & 4°C with accuracy of ±1°C
along with setting range of -20°C to +40°C
Calibration Must be calibrated for speed & Time for
various component preparation
Special eyelet window for quick calibration
Imbalance Contents in opposing cups must be equal in
weight to improve centrifuge efficiency
Imbalance Cutoff with indication to ensure
component quality as well as safety

° AABB Technical Manual. 15th edition; 2005: Table 7.4.1. Centrifugation for Component Preparation; pg.827

° Transfusion Medicine Technical Manual OCHS, 2nd edition: 2003, Blood component preparation B Use, pg. 195′

Need of Centrifuge Optimization for Quality Enhancement

Each individual centrifuge must be evaluated for the preparation of the various components.

AABB Technical Manual. 15th edition; 2005: Table 7.4.1. Centrifugation for Component Preparation; pg.827  

Centrifugation Variables : Centrifugation speeds (relative centrifugal force) and times should be standardized for each piece of equipment.

AABB Technical Manual, 15th edition; 2005; General & laboratory Methods Centrifugation Variable; pg.716

High g forces are of theoretical concern because they may damage the platelets when they are forced against the wall of the container.

AABB Technical Manual, 15th edition: 2005; Methods Section 6; Blood Collection, Storage, and Component Preparation; pg 817

REMI Provides Application Support for Quality Validation & Standard Results