Stability Chambers
| Laboratory Products | Stability Chambers
REMI Stability Chambers are designed as per ICH Guidelines for precise monitoring of temperature and humidity conditions required for long term, intermediate & accelerated stability studies at chemical, pharmaceutical, agrochemical manufacturing facilities. These chambers are manufactured to comply with GMP requirements.
- Fulfills storage conditions of 25°C – 60% RH, 40°C – 75% M RH, 30°C – 65% RH
- Precise monitoring of temperature and humidity conditions
- Unique design of thermal barrier for better energy efficiency
- Machine filled CFC free PUF insulation to eliminate void pockets
- Powerful fan motor for forced air circulation to maintain uniform conditions inside chamber within permissible limit Heating by long life SS tubular heaters
- Hermetically sealed Compressor with CFC free refrigerant
- Chamber calibration port on side
- Adjustable trays
- Output for remote alarm
- Microprocessor controller with LCD display & capacitance type sensor
- 4 ” attractive LCD display for direct display of humidity & temperature
- Auto tuning of controller
- Battery Backup for Controller
- Settable High / Low alarm points
- Centronic interface to record temperature, Humidity, Date & Time by attaching Dot Matrix printer with adjustable print interval
- REMI “Datasoft” data acquisition software complying to 21 CFR part 11 as per USFDA Guidelines (Optional)
- Power failure and resumption recorded with date and time
- PC communication through RS – 485 with data acquisition software
- High temperature safety cut off & alarms for high/low set parameters
- Low water level alarm and power cutoff of boiler tank heater
- Time delay for compressor switch on
- Overload cut off relay for compressor
- Rounded inner chamber for easy cleaning
- Auto door closer
- Attractive door profile
- Data logger for 8 point (4+4) temperature / humidity recording with sensor, printer interface and software (Optional)
- Weight carrying capacity of each tray : 30 Kg Uniformly distributed
- Smart Handle with LED door Closure indication
- Optional
- Touch screen display ( TFT ) with additional features(Details overleaf)
- Validation protocol with IQ, OQ, PQ Documentation
- Port for calibration probes / safety back up
LCD Display
- 4 Inch attractive LCD display
- Intelligent Controller helps maintain temperature in case of sensor failure
- Battery Backup for Temperature Controller
- Auto tuning of controller
- Self-Diagnosis for errors
- Histogram format of 24 hours temperature recording
- Settable High / low alarm points
- Memory for storage of upto 1000 records
- Centronic interface to record Temperature, Date, & Time by attaching Dot Matrix printer with adjustable print interval
- Alarm on equipment / output for remote alarm
Touch Screen Display
- 7 inch big colour touch screen
- PLC (Programable logic controller)
- Memory of 1500 records
- Centronic interface to record Temperature, Date & Time by attaching Dot Matrix printer with adjustable print interval
- Ethernet port
- Battery backup in case of power failure for display, alarms & off-line data logging (optional)
- Programmable cyclic timer for controlling illumination conditions
- Remi “Datasoft” Data acquisition software complying to 21 CFR part 11 as per USFDA Guidelines (optional)
REMI ‘Data Soft’
- Multiple level alpha numerical password with password ageing
- Secured audit trail report
- Alarm report with mention of alarm condition
- Mean Kinetic Temperature able to be calculated for any days & alarm deviation report
- Reports in graphical and tabular form
- Data Stored in PC server with Lan connectivity and can be monitored on PC with password authentication
- Multi chamber data acquisition on single software
- Communication through RS 485 / TCP IP
- GSM Mobile Alert for six Numbers
- Power failure and resumption recorded with date and time
Optional Accessories
To be Ordered Separately
- Data logger for 8/16 Point temperature & Humidity recording with sensor
printer interface and software (Optional) - Software as per 21 CFR Part 11 requirements
- PLC Based Control System with HMI Touch Screen
- Biometrics Magnetic Door Access for Secured Door opening with software to maintain log
- Validation (mapping) consist of one cycle for temperature at any one set point for 24 hours on empty and loaded condition
- Standby Refrigeration System
- Standby Humidity System
- GSM alert through SIM card
Technical Data | |||||||
SC-6 Plus | SC-10 Plus | SC-12 Plus | SC-16 Plus | SC-19 Plus | SC-28 Plus | SC-35 Plus | |
---|---|---|---|---|---|---|---|
internal Volume (Liters) | 200 | 280 | 370 | 450 | 540 | 800 | 1000 |
Number of tray | 2 | 3 | 4 | 5 | 6 | 6 | 7 |
Internal Dimensions W x D x H (mm) | 660 x 655 x 450 | 660 x 655 x 650 | 660 x 655 x 850 | 660 x 655 x 1050 | 660 x 655 x 1250 | 800 x 800 x 1250 | 800 x 800 x 1580 |
External Dimensions W x D x H (mm) | 800 x 1095 x 1020 | 800 x 1095 x 1220 | 800 x 1095 x 1420 | 800 x 1095 x 1620 | 800 x 1095 x 1820 | 960 x 1370 x 1950 | 960 x 1370 x 2270 |
Temperature Range & Accuracy | 10°C to 60°C, ± 1°C | ||||||
Humidity Range (Rh) & Accuracy | 40% to 95%, ± 3% | ||||||
Temperature & Humidity control | Microprocessor with Capacitance type sensor | ||||||
Display | 4″ LCD Screen, Large size Display for ease of reading | ||||||
Power Failure Alarm | Visual Alarm | ||||||
Door Open Alarm | Audio Visual Alarm in case door open for over one minute | ||||||
Humidity/Temperature Variation Alarm | Set Humidity ± 3% Rh, Set Temperature ± 1°C, Audio Visual Alarm | ||||||
Illumination | 8 Watts Fluorescent Tube | ||||||
Internal Body Material | Stainless Steel – 316 grade | ||||||
External Body Materia | Stainless Steel – 304 grade | ||||||
Insulation | 80 mm minimum for Body & 80 mm for Door, CFC free polyurethane foam | ||||||
Noise Level | Less Than 65 db(A) | ||||||
Recommended Voltage Stabilizer | VS-04 | VS-07 | VS-07 |
Supply: 220-240 Volts, 50 Hz Single Phase.